Getting Started
6.
Choose whether the protocol should be
None
or
Terminal
. A communication
session run by a human at a keyboard typically uses a terminal protocol.
An automated session might handle its own communications without using
such a protocol.
7.
If required, change the port number by rotating multipurpose knob
a.
8.
If required, press
OK
to set the new port number.
9.
After setting up the socket server parameters, you can now have the computer
“talk” to the oscilloscope. If you are running an MS Windows PC, you could
run its default client with its command-like interface. One way to do this is
by typing “
Telnet
” in the Run window. The Telnet window will open on
the PC.
NOTE.
On MS Windows 7, you must
fi
rst enable Telnet in order for it to work.
10.
Start a terminal session between your computer and your oscilloscope by
typing in an open command with the oscilloscope LAN address and port
#. You can obtain the LAN address by pushing the
LAN
panel to view the
resulting LAN setting panel. You can obtain the port # by tapping the
Socket
Server
panel and viewing the
Port
item.
For example, if the oscilloscope IP address was
123.45.67.89
and the port
# was the default of
4000
, you could open a session by writing into the MS
Windows Telnet screen:
open 123.45.67.89 4000
The oscilloscope will send a help screen to the computer when it is done
connecting.
11.
You can now type in a standard query, as found in the programmer manual,
such as
*idn?
The Telnet session window will respond by displaying a character string
describing your instrument. You can type in more queries and view more
results on this Telnet session window. You can
fi
nd the syntax for relevant
queries and related status codes in other sections of this manual.
NOTE.
Do not use the computer’s backspace key during an MS Windows' Telnet
session with the oscilloscope.
Socket Server Terminal Protocol Mode Commands.
Following are Tektronix
Instrument Control Terminal Session Control commands:
!t <timeout> : set the response timeout in milliseconds.
!d : send device clear to the instrument.
1-6
MSO54, MSO56, MSO58, MSO58LP Programmer
Содержание MSO54
Страница 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Страница 2: ......
Страница 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Страница 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Страница 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 996: ...Appendix D Waveform Transfer WFMOutpre and CURVe Query Examples D 18 MSO54 MSO56 MSO58 MSO58LP Programmer ...