Commands listed in alphabetical order
POWer:POWer<x>:REFLevels:ABSolute:FALLMid
This command sets or queries the falling edge for mid reference level in absolute
units for the speci
fi
ed power measurement badge. The power measurement badge
is speci
fi
ed by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:REFLevels:ABSolute:FALLMid
<NR1>
POWer:POWer<x>:REFLevels:ABSolute:FALLMid?
Related Commands
POWer:POWer<x>:REFLevels:METHod
,
Arguments
<NR1>
ranges from -40000 to 40000
Examples
POWer:POWer1:REFLevels:ABSolute:FALLMid 0
sets the mid reference
level for falling edge as 0 volt.
POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis
This command sets or queries the absolute hysteresis value for the speci
fi
ed power
measurement badge. The power measurement badge is speci
fi
ed by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis
<NR1>
POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?
Related Commands
POWer:POWer<x>:REFLevels:METHod
Arguments
<NR1>
ranges from 0.0000005 to 10
MSO54, MSO56, MSO58, MSO58LP Programmer
2-553
Содержание MSO54
Страница 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Страница 2: ......
Страница 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Страница 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Страница 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 996: ...Appendix D Waveform Transfer WFMOutpre and CURVe Query Examples D 18 MSO54 MSO56 MSO58 MSO58LP Programmer ...