Adjustment/Verification Fixture Adjustment
5- 12
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
The following procedure checks the DC power supply. Refer to Figure 5--3 for
test point locations.
Parameter tested
Power supply
Equipment
Required
DMM with test leads (item 9)
Prerequisites
Warm-up time: 30 minutes for adjustment/verification fixture and test
equipment
2.
Connect the DMM (--) lead to J8 (V
EE
).
3.
Connect the DMM (+) lead to J10 (V
DD
) and check for a voltage reading of
+5 V
±
0.10 V.
4.
Connect the DMM negative (--) lead to J9 (GND).
5.
Connect the DMM positive (+) lead to J10 (V
DD
) and adjust R38 for a
voltage reading of +2.00 V
±
10 mV.
Power Supply Checks