Logic Analyzer Module Adjustment Procedures
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
5- 7
Perform the following steps to run the SELF_CAL routine. Before beginning
this procedure, be sure that no active signals are applied to the LA module.
Self calibration can fail if signals are applied to the probe during the procedure.
Prerequisites
Warm-up time: 30 minutes
Power-up diagnostics pass
1.
Ensure that the instrument has had a 30-minute warm up before attempting
the self calibration, and that the logic analyzer application is running.
2.
Disconnect any probes connected to the LA module.
3.
Select Calibration and Diagnostics from the System menu.
4.
Select the Self Calibration tab page.
5.
Select the LA module.
6.
Click the Run button to start the self calibration.
The self calibration takes several minutes to complete, depending on the
number of channels in the module. Upon successfully completing the self
calibration, the module status changes from Running to Calibrated, and the
Date and Time field is set to the present.
LA Adjustment
Procedure 1:
SELF_CAL
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......