Operating Information
2- 8
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
NOTE
. Performing the self calibration does not guarantee that all logic analyzer
module parameters operate within limits. Operation within limits is achieved by
performing the
Adjustment Procedures.
Proper operation may be confirmed by
performing the performance verification procedures in this same manual.
When to Perform the Self Calibration.
You can run the self calibration at any time
during normal operation. To maintain measurement accuracy, perform the
self calibration if more than one year has elapsed since the last self calibration.
You can check the status of the self calibration in the Calibration and Diagnostics
property sheet.
If the logic analyzer module loses power during the self calibration, rerun the
self calibration following the next power-on. The self calibration data generated
before power was interrupted must be replaced with a complete set of new data.
For best results, always perform the self calibration after at least a 30 minute
warm-up.
Running the Self Calibration.
The logic analyzer module may require several min-
utes to run the self calibration depending on the number of channels. Select
Calibration and Diagnostics property sheet from the System menu. Select the
Self Calibration tab page and select the logic analyzer module. Click on the Run
button to start the self calibration. Upon completing the self calibration the logic
analyzer module menu selection changes from Running to Calibrated.
Self Calibration for Merged Modules
If you intend to merge modules, perform the
self calibration on each individual module first, then perform a self calibration on
the merged module set.
You can run the self calibration at any time during normal operation. To maintain
measurement accuracy, perform the self calibration if more than one year has
elapsed since the last self calibration.
You can check the status of the self calibration in the Calibration and Diagnostics
property sheet.
If the logic analyzer module loses power during the self calibration, rerun the self
calibration following the next power-on. The self calibration data generated
before power was interrupted must be replaced with a complete set of new data.
For best results, always perform the self calibration after at least a 30 minute
warm-up.
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......