Performance Verification: Ajustment/Verification Fixture
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
4- 29
Use the following procedure to verify the internal clock frequency.
Parameter tested
Internal clock frequency
Equipment
required
Frequency counter (item 8)
Prerequisites
Warm-up time: 30 minutes for adjustment/verification fixture and test
equipment
1.
Set the jumper positions as listed in the table.
Jumper
Jumper name
Jumper setting
J13
P6434 Deskew/probe setup
3-5 connected
J15
Clock selection
INT
J16
Deskew
2-3 and 4-5 connected
J17
Setup/hold select
HOLD
J19
Clock polarity select
Removed
2.
Using the frequency counter, measure the oscillator frequency at pin J1-2.
The frequency should match what is listed on the Calibration Data report.
3.
Record this measurement on the Calibration Data Report.
Verify Internal Clock
Frequency
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......