![Tektronix LTA7P Series Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035087.webp)
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
5- 1
Logic Analyzer Module Adjustment Procedures
This chapter contains procedures which use the performance verification and
adjustment software to adjust the TLA7Nx, TLA7Px, and TLA7Qx logic
analyzer modules to within factory specifications. The performance verification
and adjustment software contains instructions and control programs for adjusting
the instrument The software describes test equipment connections and settings,
selects setup parameters, and loads calibration constants into memory.
This chapter also contains adjustment procedures for the adjustment/verification
fixture which begin on page 5--11.
These procedures adjust the LA Module for conformance with the warranted
characteristics listed in the
Specifications
chapter of this manual.
Adjustments should be done after repair of the module or when performance
verification tests have failed.
Prerequisites
These procedures ask for the serial number of the instrument under test. Before
installing the modules in the mainframe, record the serial number of the LA
module.
You can also access the module serial number through the logic analyzer
application. In the application, go to the System menu, select System Properties,
and click on the LA instrument tab. You must quit the logic analyzer application
before continuing with the performance verification and adjustment software
procedures.
Only trained service technicians should perform this procedure after meeting the
following requirements:
H
When multiple LA modules of the same model number are installed in the
mainframe, the performance verification and adjustment software will
address only the module in the highest-numbered slot. If you are testing a
TLA7Q4 module for example, move it to a higher slot number than all of the
other TLA7Q4 modules in the mainframe. This avoids unnecessary module
warm-up time.
H
When adjusting merged modules using the same type of individual modules,
the individual modules must be physically separated.
H
The logic analyzer application must not be running.
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......