![Tektronix LTA7P Series Скачать руководство пользователя страница 110](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035110.webp)
Maintenance
6- 2
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
H
Nothing capable of generating or holding a static charge should be allowed
on the work surface.
H
Handle circuit boards by the edges when possible.
H
Do not slide the modules over any surface.
Inspection and Cleaning
Inspection and cleaning are done as preventive maintenance. Preventive
maintenance, when done regularly, may prevent malfunctions and enhance
reliability.
Preventive maintenance consists of visually inspecting and cleaning the logic
analyzer module, and using general care when operating it. How often to perform
maintenance depends on the severity of the environment in which the logic
analyzer module is used. A proper time to perform preventive maintenance is
just before performing adjustments on the module.
The side cover keeps dust out of the module and should be in place during
normal operation.
Inspect and clean the logic analyzer module as often as operating conditions
require. Collection of dirt on internal components can cause them to overheat
and breakdown. Dirt acts as an insulating blanket, preventing efficient heat
dissipation. Dirt also provides an electrical conduction path that can cause
failures, especially under high-humidity conditions.
CAUTION.
Avoid using chemical cleaning agents that might damage the plastics
and external labels used in the logic analyzer module.
Use a cloth dampened with water to clean external surfaces. To prevent damage
to electrical components from moisture during external cleaning, use only
enough liquid to dampen the cloth or applicator.
Use a 75% isopropyl alcohol solution to clean internal surfaces and rinse with
deionized water. Before using any other type of cleaner, consult your Tektronix
Service Center or representative.
Inspect the outside of the module for damage, wear, and missing parts. Use
Table 6--1 as a guide. Instruments that appear to have been dropped or otherwise
abused should be checked thoroughly to verify correct operation and perform-
ance.
General Care
Inspection and Cleaning
Procedures
Exterior Inspection
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......