![Tektronix LTA7P Series Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035080.webp)
Performance Verification: Ajustment/Verification Fixture
4- 32
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
Use the following procedure to verify the hold time.
Parameter tested
Hold time
Equipment
required
Oscilloscope (item 3)
2 DSO probes (item 4)
Custom probe tip adapter (see Figure 4--10, page 4--26)
Prerequisites
Warm-up time: 30 minutes, adjustment/verification fixture and test
equipment
1.
Set the jumper positions as listed in the table.
Jumper
Jumper name
Jumper setting
J13
P6434 Deskew/probe setup
1-3 and 5-6 connected
J15
Clock selection
INT
J16
Deskew
1-2 and 4-5 connected
J17
Setup/hold select
HOLD
J19
Clock polarity select
Disconnected
2.
Set up the oscilloscope by pressing Setup, Factory Setup, and then OK
Confirm Factory Init to return the oscilloscope to default conditions.
3.
Set up the oscilloscope as listed below.
a.
Set up the CH1 Vertical menu as follows:
H
Coupling
DC/50
Ω
H
Fine Scale
200 mV/div
H
Position
--3.00 div
CH1
H
Position
--3.32 div
CH2
b.
Set up the Horizontal menu as follows:
H
Time Base
Main
H
Record Length
5000
H
Horizontal Scale
Main Scale @ 2 ns/div
Verify Hold Clock Output
Timing
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......