![Tektronix LTA7P Series Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035048.webp)
Theory of Operation
3- 4
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
A merged module consists of a Master module and a Slave module connected
together by a merge cable connector and by signals on the local bus of the
mainframe backplane. The local bus sends the system clock of the Master
module to the Slave module. The two merged modules must be located in
adjacent slots.
The merge Cable Connector passes 26 signals between the two modules (16
trigger event signals, two storage control signals, foru valid sample clock iden
tification signals, and four data-login control signals).
Merged Modules