![Tektronix LTA7P Series Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035059.webp)
Performance Verification: Logic Analyzer Module
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
4- 11
H
The LA module must be installed in a mainframe, operating for at least
30 minutes, and operating at an ambient temperature b20
_
C
and +30
_
C.
H
The LA module must have been last adjusted at an ambient temperature
b20
_
C and +30
_
C.
H
The logic analyzer must be in an operating environment within the limits
described in the
Specifications
section of the
Tektronix Logic Analyzer
Family User Manual
.
H
When verifying the performance of merged modules consisting of different
types of individual modules, the merged module can be tested without
separation. The performance verification and adjustment software runs
independent of the logic analyzer application and does not recognize
configuration settings. It is unnecessary to unmerge modules through the
logic analyzer application before performing these procedures.
When using the performance verification and adjustment software, you will
connect external test equipment to the LA module in response to prompts on the
screen. You will connect the test signals and then instruct the program to
continue. The performance verification and adjustment software automatically
selects the module settings and determines the results of each test.
The results of the tests are recorded in a temporary file and are available upon
test completion for completing test records for certification. To obtain partial test
information you can also run individual tests or selected groups.
NOTE
. The SELF_CAL test must run successfully before the other tests are
performed. The remaining tests can then be performed in any order.
Before testing an instrument following repair, you must first complete the
adjustment procedure.
The performance verification and adjustment software contains the tests shown
in Table 4--3. Each test verifies one or more parameters. All of the tests check
characteristics that are designated as checked (
n
) in the
Specifications
chapter
.
By running a full PV sequence, you will verify the performance of the LA
Module.
Procedure Overview
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......