![Tektronix LTA7P Series Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035063.webp)
Performance Verification: Logic Analyzer Module
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
4- 15
The results of all tests can be stored in a file on the hard disk. You can view the
test results, print the test results to a printer, or save the test results in another file
on the hard disk. The software stores the test results in a file containing the
module name and serial number (for example, TLA7Q4.B020123). The file is
located under the following path: C:\Tekcats\Rpt.
NOTE
. If you want to save the content of the Report file, you must rename or
copy the Report file using the Windows file utilities such as Explorer.
The Report file will be overwritten the next time you run the performance
verification and adjustment software and print or view a new set of test data.
After completing a full or partial test sequence (or just before you exit the
program) you can generate the test data and write it to a file. You have the option
of printing the file, viewing the file on screen, or transferring the file to another
directory or host computer.
You can print the test data directly from the program. Ensure that a printer is
connected to your logic analyzer and follow the on-screen instructions to print
the test results.
If a printer is not available, you can view the test results directly from the screen,
or you can copy the test results to a different file or folder/host computer for
future use.
The performance verification and adjustment software creates data log files (.dlf
files) that store program data. The .dlf files are used by the performance
verification and adjustment software to generate the view data and print out
options. Each .dlf file is identified by the product serial number; for example,
B010100.dlf. The files are stored in the Tekcats folder under each TLA7xx
folder. To conserve disk space, you must occasionally delete the .dlf files.
If any tests fail, use the following steps to troubleshoot the problems:
1.
Check all test equipment for improper or loose connections.
2.
Check that all test equipment is powered on and has the proper warm-up
time.
3.
If you are using the adjustment/verification fixture, verify the LED is
lighted, the jumper positions match the on-screen instructions, and the
external connections are correct. (See Figure 4--4 for jumper locations.)
4.
Rerun mainframe or module diagnostics and module adjustment.
5.
Run the tests a second time to verify the failure.
Obtaining Test Results
Field Adjust/PV Software
Housekeeping
Troubleshooting
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......