Performance Verification: Ajustment/Verification Fixture
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
4- 33
c.
Set up the Trigger menu as follows:
H
Source
CH2
H
Coupling
DC
H
Slope
+
H
Level
700 mV
H
Mode
Normal
d.
Set up the Measure menu as follows:
H
Select Measurement for CH1
Measure
H
Gating
On
e.
Set up the Cursor menu as follows:
H
Function
V Bars
f.
Set up the Acquire menu as follows:
H
Acquisition Mode
Average 90
H
Repetitive Signal
On
4.
Connect the CH2 probe of the oscilloscope to J1-2 on the adjustment/verifi-
cation fixture.
5.
Select Deskew from the Vertical menu.
NOTE
. Observe proper polarity when performing the following steps.
See Figure 4--9 on page 4--23 to identify the ground side of the components being
measured.
6.
Connect the CH1 probe of the oscilloscope to the custom probe tip adapter
and measure the signal across R67.
7.
Begin the acquisition by pressing the Run/Stop button.
8.
Use the large knob at the upper right of the panel to adjust delay until the
leading edges of the two waveforms coincide and the displayed value of
CH1-CH2 delay is averaged around 0.
9.
Connect the CH1 probe to R127. Verify the CH1-CH2 delay is averaged
around 0 ns
±
100 ps.
Содержание LTA7P Series
Страница 4: ......
Страница 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 86: ......
Страница 96: ...Logic Analyzer Module Adjustment Procedures 5 10 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Страница 171: ......
Страница 172: ......