Performance Verification
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
67
Recovered Clock to Data Random Jitter
Equipment
required
One sampling oscilloscope with SD22 sampling head
Semi-rigid cable for SD22 sampling head
One coaxial cables
One optical cable
GBIC, 1.25 GB/s
Prerequisites
See page 43
NOTE
. Make sure that the optical connector ends of both the fiber and the input
to the GTS1250 are well cleaned before performing this step.
1. Connect your wrist strap to the antistatic connector on the front of your
oscilloscope.
2. Install a GBIC in the GBIC Test System, and connect the optical fiber
between the Tx and Rx connectors of the GBIC.
NOTE
. The longer wavelengths of 1310 nm and especially 1550 nm are sensitive
to loss in fiber due to bending of the fiber. The fiber bend radius of the GTS1250
fiber input should lay with >1.5 inch bend radius along the fiber’s entire length.
This precaution must be maintained throughout the entire performance
verification procedure.
3. Connect an SMA cable between the RECOV. DATA output of the GBIC Test
System and the oscilloscope Ext trigger input as shown in Figure 30.
4. Connect an SMA cable between the RECOV CLK output of the GBIC Test
System and the SD-22 input of the oscilloscope as shown in Figure 30.
Select K28.5
Tx pattern
GBIC
Tx
Rx
Tektronix
1180C or
CSA80C
Oscilloscope
SD-22
Ext Trig input
RECOV. DATA
RECOV. CLK
GBIC Test
System
Figure 30: Recovered Clock to Recovered Data Random Jitter Test
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com