Fuse Removal and Replacement
74
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
Screws (8)
Top cover
Figure 33: Removing the top cover
Refer to Figure 34 for the following steps:
4. Disconnect the power supply cable from the power supply.
5. Disconnect the TTL remote cable from the main logic board.
6. Carefully lift the clear plastic shield located over the power supply.
7. Replace the fuse cartridge on the Power Supply board with the fuse listed in
the Replaceable Parts section beginning on page 77.
8. Reposition the power supply shield back over the Power Supply board.
9. Connect the power supply cable to the power supply.
10. Connect the TTL remote cable to the main logic board.
11. Place the top cover on the bottom cover assembly. Check that the plastic
power supply shield is tucked inside the bottom cover assembly.
12. Insert the (8) machine screws through the top cover into the bottom cover
assembly. Tighten the screws using a torque driver set to 4 ft-lb.
13. If other cables were removed, reinstall them using Table 22 on page 76 as a
guide.