Specifications
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
37
Table 14: Environmental Characteristics
Specification
Description
Temperature, non-operating
Class 6 (–55
_
C to +75
_
C)
The environmental exposure is the procedure stated in Tektronix Design Standard
062-2847-00 for Class 6 equipment.
Temperature, operating
10
_
C to +40
_
C (50 to 104
_
F)
This operating condition is guaranteed for the instrument itself independent of the GBIC
installed.
This procedure conforms to the temperature portion of the test in MIL-T-28800E for
Type III, Class 6 equipment, using the limits above.-
Humidity, non-operating
0% to 75% relative humidity from +30
_
C to +60
_
C.
Humidity, operating
0% to 75% relative humidity from +10
_
C to +40
_
C.
Random vibration, non-operating
2.46 g RMS, from 5 to 500 Hz, 10 minutes each axis.
Tektronix Standard 062-2858-00 rev. B modified, Random Vibration, Tektronix Class 3
with power spectral density break-points as follows:
Constant 0.020 g
2
/Hz from 5 Hz to 100 Hz, then falls at –3 dB/octave from 100 to
200 Hz, then constant 0.010 g
2
/Hz from 200 to 350 Hz, then falls at –3 dB/octave from
350 to 500 Hz to 0.007 g
2
/Hz @ 500 Hz.
Tested both with and without an IBM GBIC-1250 GBIC module installed.
Random vibration, operating
0.31 g RMS, from 5 to 500 Hz, 10 minutes each axis.
Tektronix Standard 062-2858-00 rev. B, Random Vibration, Tektronix Class 3
Tested both with and without an IBM GBIC-1250 GBIC module installed.
Packaged product vibration and shock
The packaged product qualifies under the Distribution Cycle 1 Assurance Level II for
packaged products 0 to 20 lbs. Test 2 for Warehouse and Vehicle Stacking (compression)
is omitted.
Tektronix standard 062-2858-00, Rev. B.
Tested both with and without an IBM GBIC-1250 GBIC module installed.
Altitude, operating and nonoperating
Operating:
3000 m (9847 ft)
Nonoperating:
12190 m (40,000 ft)
MIL-T-2880E modified for more exposure time and more severe test levels than required
for any Type III equipment regardless of Class.
Weight, shipping
5.22 kg (11.5 lbs.), including standard accessories
2
The humidity limits are derived from a psychrometric chart using a maximum wet bulb temperature of 29
_
C. The highest
RH (20%) at the maximum temperature (+60
_
C) diverges from the psychrometric chart (which would indicate 6%). The
20% number is a practical limit. Test chambers cannot be set below this number.