Performance Verification
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
59
f.
Set the Vertical Offset to 0.0 V/div.
g. Set the Horizontal Size to 500 ps/div.
h. Set the Main Position to any value less than 50 ns.
i.
Set the oscilloscope record length to 512 (maximum) points.
j.
Set the oscilloscope horizontal positioning mode to expand about the
CENTER.
k. Select the WAVEFORM menu and then the Sampling Head Fnc’s pop-up
menu. Set Smoothing to Off.
l.
Select the UTILITY menu and then the Instrument Options pop-up
menu. Set Vectored Trace to Off.
m. Use the oscilloscopes controls, place the resulting Tx Data eye pattern on
the display such that the crossing of the eye is at the very center of the
display.
n. Adjust the oscilloscope controls to expand the displayed waveform such
that the vertical scale is 20 mV per division and the horizontal scale is
10 ps per division. While expanding the waveform, readjust the
horizontal and vertical controls to maintain the crossing of the eye at the
center of the display.
o. Set the oscilloscopes DISPLAY MODE to generate a horizontal
histogram. Select and then adjust the VERTICAL LIMITS to create a
rectangular histogram box 2 mV in vertical height and placed such that it
intersects across the crossing of the eye (see Figure 26).
6. Perform the test with the following sequence on the oscilloscope:
a. Expand the Persistence/Histogram menu. Set the STOP on N waveforms
mode to 100 waveforms. Select the STOP on N waveforms mode (this
will cause the oscilloscope to start a new persistence acquisition in
which the oscilloscope stops acquiring after 100 waveforms are
acquired).
NOTE
. If this fast method yields a result which might indicate marginal
deterministic jitter performance, use the ten edge method for determining the
true deterministic jitter of the K28.5 pattern.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com