Performance Verification
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
47
4. Connect SMA cables between the Tx DATA and Tx DATA outputs of the
GBIC Test System and the SD-22 inputs of the oscilloscope as shown in
Figure 17.
5. Set the GBIC Test System Transmit pattern to D21.5.
Select D21.5
Tx pattern
GBIC
Tx
Rx
Tektronix
1180C or
CSA80C
Oscilloscope
Sync
GTS1250
Tx DATA
Tx DATA
SD-22
Ext Trig input
Figure 17: Tx Data levels (swing)
6. Perform the initial setup of the oscilloscope using the following steps:
a. Initialize the oscilloscope to default settings using the Initialize selector
which appears in the Utility major menu.
b. From the Trigger major menu, set the trigger source to external.
c.
Set the trigger level to 0.0 V; set the trigger polarity to positive edge.
d. Display both SD-22 channels.
e.
Set the Vertical Size of both channels to 100 mV/div.
f.
Set the Horizontal Size to 500 ps/div.
g. Select the WAVEFORM menu, and select the Horizontal Desc. menu.
h. Set the oscilloscope record length to 512 points.
i.
Set the oscilloscope to 32 averages on both channels.
j.
Select the DISPLAY MODES menu.
k. Select the Persist/Histograms pop-up menu, select the Normal menu, and
then select Exit.
NOTE
. For increased accuracy, you may perform a LOOP Gain adjustment of the
SD22 head on the oscilloscope. If you need instructions, refer to the oscilloscope
and SD22 user manuals.
7. Perform the test with the following sequence on the oscilloscope: