Getting Started
1–8
DAS System User Manual
The mainframe and terminal should connect to a power source that meets the
requirements stated in this section. The mainframe must be supplied with a properly
rated power cord. The mainframe power requirements depend on the number and
type of modules installed. To determine which power source your mainframe is set
to use, check the line voltage indicator on the rear panel of the mainframe.
Terminal Power Requirements.
The color terminal requires a power source that
allows one-cycle surge currents of less than 30 A.
Mainframe Power Requirements.
The Master mainframe and Expansion main-
frames support different numbers of acquisition and pattern generation cards
based on the type of power cord/supply being used with each mainframe. The
standard 12 A power cord should be used for mainframe configurations drawing
less than 400 watts (12 A/115 V) and for Expansion mainframes drawing less
than 475 watts. When the configuration of cards in the mainframe requires more
power, a higher-current power cord must be used (refer to Tables 1–1 and 1–2). If
installed cards require more than 700 watts (750 in Expansion mainframes), the
DAS will automatically shut off. If the power requirements of the cards installed in
the mainframe exceed the power cord and power supply configuration, the DAS
also automatically shuts off. Contact your Tektronix field service representative for
assistance in determining the proper power supply or power cord configuration.
NOTE
. If you install the 15 A power cord Option 1A, be sure to affix the label
supplied with the power cord. This label should be positioned directly to the
right of the power cord connector on the rear panel; the label should be placed
over the existing description that reads: 1900 VA MAX., FREQ. 48-63 Hz.
The maximum continuous current that each Master or Expansion mainframe can
draw is 15 A; the one-cycle surge current is 25 A nominal. Total line current for a
mainframe with three Expansion mainframes can be up to 60 A @ 115 V line
selection (40A @ 230 V). For a system using the maximum of four mainframes, the
power source must be capable of supplying the maximum amount of line current.
When the system is connected to a 230 V power source, the 230 V power cord
must be used. The DAS is specified to draw no more than 10 A when set for
230 V operation. When the DAS is connected to a three-phase power source, the
mainframe is specified to draw no more than 8 A on any phase; the three phase
power cord is permanently attached to the power supply. If it becomes necessary to
upgrade your power supply, contact your nearest Tektronix field service center.
Power Requirements
Содержание DAS 9221 Option 04
Страница 4: ......
Страница 6: ......
Страница 16: ...Table of Contents x DAS System User Manual ...
Страница 20: ...General Safety Summary xiv DAS System User Manual ...
Страница 23: ......
Страница 48: ......
Страница 81: ......
Страница 156: ......
Страница 166: ...Appendix B Accessories B 4 DAS System User Manual ...
Страница 178: ...Appendix C Error Messages C 12 DAS System User Manual ...
Страница 238: ...Appendix F User Service F 26 DAS System User Manual ...
Страница 249: ......
Страница 266: ......
Страница 275: ......
Страница 276: ......