Acquiring Waveforms
CSA8000B & TDS8000B User Manual
3- 11
With an input signal connected, use the procedure that follows to autoset based
on the characteristics of the input signal. Autoset operates on the selected
channel only.
Overview
To autoset
Control elements and resources
Prerequisites 1.
The instrument must be installed with sampling modules
in place. Signals must be connected to channels. A
triggering source must be provided.
2.
At least one channel must be turned on (its front-panel
button lighted).
See the sampling-module user manuals for help with
installing sampling modules. See page 3--48 in this
manual for trigger setup information.
Execute 3.
Push the
Autoset
button to to execute an autoset on the
selected waveform.
If you use Autoset when one or more channels are
displayed, the instrument uses the selected channel for
horizontal scaling. Vertically, all channels in use are
individually scaled.
Note.
Autoset can execute on live waveforms (either
channel or math) in the Main time base.
To Autoset the Instrument
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......