Using Masks, Histograms, and Waveform Databases
CSA8000B & TDS8000B User Manual
3- 145
H
Masks are saved with setups, so you can save sets of masks by defining
them, and then storing the instrument setup. Displayed masks are overwrit-
ten when you recall a stored setup, select a standard mask, or initialize the
instrument.
Use the procedure that follow to set up the instrument to mask test a waveform
against a mask standard or user-defined mask set.
Overview
To mask test a waveform
Related control elements & resources
Prerequisites 1.
The instrument must have at least one waveform turned
on.
See
Displaying Waveforms
on page 3--53 for
information on displaying waveforms.
Access the Mask
Setup dialog box
2.
Select
Mask
from the Setup menu to display the Mask
Setup dialog box.
To Mask Test a Waveform
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......