Incoming Inspection
CSA8000B & TDS8000B User Manual
1- 23
CSA8000/TDS8000
SMA cable from DC calibration
output to 80E00 C3 input
Figure 1- 7: Hookup for electrical functional tests
4.
Set the DC CALIBRATOR OUTPUT:
a.
Push the Vertical
MENU
front-panel button. This displays the Vert
Setup dialog box.
NOTE
.
When an optical module is installed, the optical setup dialog box displays
by default. Click the Basic button to display the basic dialog box.
b.
Enter a level of
200 mV
in the DC CAL box.
c.
Push the Vertical
MENU
front-panel button again to dismiss the Vert
Setup dialog box.
5.
Select the channel to test:
Push the channel button for the channel you want
to test. The button lights and the channel display comes on. See Figure 1--8.
Channel
buttons
Figure 1- 8: Channel button location
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......