Appendix B: Automatic Measurements Reference
CSA8000B & TDS8000B User Manual
B- 21
Table B- 4: RZ Measurements - Amplitude (cont.)
Name
Definition
RZ Max
The maximum vertical value of the waveform that is sampled within the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
When this measurement is turned on, it will automatically set the measurement system to use a
waveform database if available.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
To Optimize the
Vertical Resolution
on page B--69.
RZ Mean
The arithmetic mean of the waveform that is sampled within the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
When this measurement is turned on, it will automatically set the measurement system to use a
waveform database if available. See
Use a Waveform Database
on page B--70.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
How to Optimize the
Vertical Resolution
on page B--70.
RZ Mid
The middle level between the Max and Min vertical values of the waveform that is sampled
within the measurement region.
Where Max and Min are the maximum and minimum measurements.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
When this measurement is turned on, it will automatically set the measurement system to use a
waveform database if available.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
To Optimize the
Vertical Resolution
on page B--69.
RZ Mid
=
(
Max
+
Min
)
2
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......