Incoming Inspection
CSA8000B & TDS8000B User Manual
1- 27
NOTE
. If the position knob was set to 0.000, you can confirm this in the Vertical
menu (use Basic button in the dialog box).
Control bar
Vertical offset
setting
Baseline
Vertical offset
Figure 1- 10: Optical channel verification
5.
Verify that the channel acquires in all acquisition modes:
Push the
front-panel button Acquisition
MENU
to display the Acq Setup dialog box.
Click each of the three acquisition modes and confirm that the following
statements are true:
H
Sample
mode displays an actively acquiring waveform on-screen. (Note
that there may be a small amount of noise present on the baseline level).
H
Average
mode displays an actively acquiring waveform on-screen with
any noise present reduced.
H
Envelope
mode displays an actively acquiring waveform on-screen with
the upper and lower extremes of the noise displayed.
6.
Close Acquisition setup dialog box:
Push the Acquisition
MENU
button to
close the Acq setup dialog box.
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......