Appendix B: Automatic Measurements Reference
B- 4
CSA8000B & TDS8000B User Manual
Table B- 1: Pulse Measurements — Amplitude (cont.)
Name
Definition
Gain
The amplitude gain between two waveforms. The measurement returns the ratio between the
amplitudes measured within the measurement regions of the two sources.
Where Amplitude1 and Amplitude2 are the Amplitude measurements of the two source
waveforms. See
Amplitude
on page B--2.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2) of the two sources. See
To Localize a Measurement
on
page 3--83.
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Gain
=
Amplitude1
Amplitude2
High
The top reference level for a measured waveform. Several methods can be applied to the data
sampled in the upper half of the waveform to determine the High value. You can use the
Tracking Method control to select among them. See
Tracking Methods
on page B--66. Also see
High/Low Tracking Method
on page 3--77.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Low
The bottom reference level for a measured waveform. Several methods can be applied to the
data sampled in the lower half of the waveform to determine the Low value, and you can use
the Tracking Method control to select among them.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Max
The largest amplitude peak of the waveform over the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Mean
The arithmetic mean of the waveform over the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......