Appendix B: Automatic Measurements Reference
B- 20
CSA8000B & TDS8000B User Manual
Table B- 4: RZ Measurements - Amplitude (cont.)
Name
Definition
RZ High
The logical 1 level of the RZ signal. The data within the Eye Aperture is sampled, a histogram is
built from the upper half of the RZ eye, and the mean of the histogram yields the High level.
The Eye Aperture is adjustable and defaults to 5% of the RZ pulse width. See
RZ Eye Aperture
Parameters
on B--62.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See
Use a Waveform Database
on page B--70.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
How to Optimize the
Vertical Resolution
on page B--70.
RZ Low
The logical 0 level of the RZ signal. The data within the Eye Aperture is sampled, a histogram is
built from the lower half of the RZ eye, and the mean of the histogram yields the Low level.
The Eye Aperture is adjustable and defaults to 5% of the RZ pulse width.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
To Optimize the
Vertical Resolution
on page B--69.
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......