Acquiring Waveforms
3- 24
CSA8000B & TDS8000B User Manual
Use the procedure that follows to set the data-acquisition mode and specify
acquisition start and stop methods. For more detailed information, display online
help when performing the procedure.
Overview
To set acquisitions modes
Control elements and resources
Prerequisites 1.
Instrument must be installed with sampling modules in
place before powering on the instrument. Instrument
must be powered up, with horizontal and vertical
controls setup. Triggering should also be set up.
See the sampling-module user manuals for sampling
module installation. See page 3--48 for trigger setup.
To select an
Acquisition mode
2.
Push the Acquisition
MENU
button to display the Acq
Setup dialog box.
Select the
Acquisition mode
Set a
sample count
3.
Click an option button to select the acquisition mode;
choose from the following modes:
H
Sample
H
Average
H
Envelope
For Average mode only, enter the number of samples to
to average in the Average box.
To Set Acquisition Modes
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......