Incoming Inspection
1- 30
CSA8000B & TDS8000B User Manual
NOTE
. At some temperatures, there may be extraneous data points after the first
half cycle when viewing the front-panel Internal Clock output (as is done in this
step). This behavior may also occur when viewing multiple cycles in TDR mode.
In both cases, this behavior is normal.
H
Rotating the Horizontal
SCALE
knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counter-
clockwise rotation contracts it, and returning the horizontal scale to
1
s/div
returns the period to about five divisions.
Leave the time base
set to 1
s/div.
H
The horizontal
POSITION
knob positions the signal left and right
on-screen when rotated.
NOTE
. The signal will not move past the minimum position setting.
Control bar
Vertical scale
setting
Horizontal
scale setting
Internal clock
signal
s
Figure 1- 13: Main time base verification
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......