Incoming Inspection
1- 34
CSA8000B & TDS8000B User Manual
7.
Push the Horizontal
MENU
button, the Mode in All Timebases must be set
to Lock to Int. 10 MHz.
8.
Verify that Triggering occurs:
Verify signal is triggered with waveform
on-screen. See Figure 1--17 on page 1--34.
Control
bar
Vertical scale
setting
Horizontal
scale setting
Internal
clock
signal
Triggered signal indicator
Figure 1- 17: Signal triggered
9.
Disable trigger:
Install 50
Ω
terminator cap to the end of the cable that is
attached to the rear-panel gated trigger BNC. See Figure 1--16 on page 1--33.
10.
Verify that the Gated Trigger functions:
Verify signal is not triggered (gate
disabled). Signal freezes on the screen above to indicate triggering has
stopped. See Figure 1--18 on page 1--35. Note the Not Trigd indication at the
top of the window.
a.
Push the
CLEAR DATA
button.
b.
Verify signal is not triggered with no waveform on-screen (see Fig-
ure 1--19 on page 1--36). Note the Not Trigd indication at the top of the
window.
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......