Appendix B: Automatic Measurements Reference
B- 50
CSA8000B & TDS8000B User Manual
Non-Return-to-Zero (NRZ) Measurements - Timing
Table B--8 topic describes each NRZ measurement in the timing category. See
Table B--7 on page B--37 for amplitude category measurements.; see Table B--9
on page B--55 for area category measurements.
Table B- 8: NRZ Measurements - Timing
Name
Definition
NRZ Bit Rate
The inverse of the time interval between two consecutive eye-crossing points. In other words, it
is the reciprocal of the Bit Time.
Where Tcross2 and Tcross1 are the mean of the histogram of the two consecutive eye
crossings. See
NRZ Crossings
on page B--64.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See
Use a Waveform Database
on page B--70.
NRZ Bit Rate
=
1
(
Tcross2
−
Tcross1)
NRZ Bit Time
NRZ Bit Time is measured as the time interval between two consecutive eye-crossing points.
NRZ Bit Time = Tcross2 – Tcross1
Where Tcross2 and Tcross1 are the mean of the histogram of the two consecutive eye
crossings.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
NRZ Crossing Time
The horizontal position of the eye crossing. Data is sampled on a horizontal slice at the eye
crossing, and the mean of the horizontal histogram returns the crossing time.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
Содержание CSA8000B Series
Страница 4: ......
Страница 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Страница 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Страница 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Страница 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Страница 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Страница 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Страница 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Страница 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Страница 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Страница 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Страница 353: ......
Страница 354: ......