Adjustment-5110
b.
Apply a
ground
connection
(short circuit) between
the junction of R885 and
C890, and a ground test point.
See
Fig. 3-1 for the junction
and ground test point
locations.
c.
Check
—for
a meter reading of +395 to +405
millivolts.
d. Disconnect the
ground connection (short circuit)
from
the
junction and ground test point.
e. Disconnect the digital voltmeter.
NOTE
If
the high-voltage
power supply is within the
specified
tolerance,
proceed with part g. If the
adjustment
is to be made, all remaining adjustments
in
this procedure could be affected and should be
performed to
verify the
accuracy of all adjustments.
An insulated
screwdriver must be used to adjust
variable
components in this instrument, especially in
the
high-voltage area, to prevent shorting voltages
to ground
and damaging the instrument.
5. Adjust
High-Voltage Power Supply
a. Press
the POWER switch to turn off the oscilloscope
and return the oscilloscope to its normal
upright position.
b.
On the
rear panel of
the instrument, remove the two
cap
nuts securing
the cover over the crt socket, then
remove the cover (a5/16-inch nutdriver may be needed to
remove the cap
nuts).
c. Set
the de
voltmeter (vom) to measure at least -4000
volts de.
Remove the insulating sleeve from
the probe tip
of the
test
lead
to be used for measuring the negative
voltage.
Connect
the voltmeter leads between a con
venient chassis ground
and the high-voltage test point.
See Fig. 3-2 for test point location. (The high-voltage lead
should be
fully
inserted through
the crt socket cover
so
that the lead
connects to the test
point without having to
hold
it by hand).
d.
Pull
the
POWER switch to turn on the oscilloscope.
e.
Check—for a meter
reading of -3400
volts, ±170
volts.
Fig.
3-2.
Location
of high-voltage test point.
f.
Adjust
—High
Volts Adjust
R275, using an insulated
screwdriver,
for a meter reading of exactly —3400 volts.
See Fig.
3-3 for adjustment location.
g.
Press
the
POWER
switch to turn off the oscilloscope
before disconnecting
the voltmeter.
h.
Disconnect the de voltmeter and replace the cover
over the crt socket, reversing the
procedure
given in part b
of this step.
6. Adjust
Intensity Range
a.
Pull the POWER switch to turn on the oscilloscope.
b.
Set the INTENSITY control fully counterclockwise.
Fig. 3-3. Locations of high-voltage,
Intensity,
and display
adjustments.
2IJ4
09
3-4
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......