Operating lnstructions-5110
5.
Adjust
the horizontal
Position control to move the
10%
point of the waveform
to the second vertical line of the
graticule.
6.
Measure the
horizontal
distance between the 10%
and
90% points. Be
sure the variable
Seconds/Div control
is in
the Cal
position.
7. Multiply the distance
measured in step 6 by the
setting
of the Seconds/Div switch.
EXAMPLE:
Assume
that the horizontal distance
between
the
10% and 90%
points is four
divisions and the
Seconds/Div switch
is set to 1
µs.
Using
the period
formula to find risetime:
Risetime
period
horizontal
Sec/Div
distance X
switch
—
(4) (1
µs)
- 4
µs
(divisions)
setting
The
risetime is 4 microseconds.
4. Connect the reference signal to
the Channel 1 input
connector
and
the comparison signal
to
the Channel 2 (or
center
plug-in)
input
connector. The reference signal
should
precede the comparison signal in time.
Use
coaxial
cables
or
probes
which
have similar time-delay
characteristics
to
connect the signal to the input connec
tors.
5.
If
the
signals are of opposite
polarity, invert the
Channel 2 (or center
plug-in) display. (Signals may
be of
opposite
polarity
due to 180° phase difference; if so,
take
this into account
in the final calculation.)
6.
Set
the
Volts/Div switches to
produce about four
divisions
of display waveform.
7.
Set
the time-base
triggering
controls for a stable
display.
Set the Seconds/Div switch
for
a sweep rate which
shows
three or
more divisions between the measurement
points,
if possible.
8.
Adjust
the vertical Position controls to bring the
measurement
points
to
the center
horizontal
reference
line.
9.
Adjust
the horizontal
Position control so the
Channel
1 (or left plug-in) waveform (reference) crosses
the center
horizontal
line at a vertical graticule line.
10. Measure the
horizontal distance between the two
measurement points
(see Fig.
1-7).
Time
Difference Measurements
When
used in conjunction with a calibrated time-base
plug-in
unit, the multi-trace
feature of the oscilloscope
permits
measurement of time difference between
two or
more
separate
events. To measure ti me difference, use the
following procedure:
1.
Set the
input coupling switches of
the amplifier
channels
to
either ac or de.
2.
Set
the Display switch on the time-base unit to either
Chop or Alt.
In general, Chop is more suitable for low-
frequency
signals. More information on determining the
mode
is
given
under Vertical Display Mode in thissection.
3. Set the vertical plug-in triggering switches
to trigger
the display
on
Channel 1 (or
left plug-in) and Channel 2 (or
center
plug-in).
Fig.
1-7. Measuring time difference between two pulses.
1-12
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......