Operating lnstructions-51
10
11. Multiply
the
measured distance
by the setting of
the
Seconds/Div switch.
EXAMPLE: Assume that the Seconds/Div
switch is set
to 50 µs and the
horizontal distance between measure
ment points is four divisions.
Using the formula:
6.
Set the
Volts/Div switches and the variable
Volts/Div
controls
so the displays are equal and about
five divisions
in
amplitude.
7.
Set
the
time-base triggering controls to obtain a
stable
display. Set the
Seconds/Div switch to a sweep rate
which
displays about one cycle of the waveform.
Time
Delay
Sec/Div
horizontal
-
switch X
distance
= (50 µs) (4) = 200 µs
setting
(divisions)
The time delay is 200 microseconds.
Multi-trace
Phase Difference
Measurement
Phase
comparison between two or more signals of
the
same
frequency can
be made using a dual-trace
plug-in or
two
single-trace
plug-ins. This method of phase difference
measurement
can be used up to the frequency limit of the
vertical
system. To make the comparison, use the
follow
ing
procedure:
1.
Set
the input coupling, switches of the amplifier
channels
to either ac or de.
2.
Set
the
Display switch on
the time-base unit to either
Chop or Alt. In
general, Chop is more suitable for low-
frequency
signals
and the
Alt position is more suitable for
high-frequency signals. More information on determining
the mode is given under Vertical Display Mode in this
section.
3.
Set the vertical plug-in
triggering
switches to trigger
the
display
on Channel 1 (or left plug-in) and Channel 2 (or
center
plug-in).
4.
Connect
the reference signal to the Channel 1 input
connector
and comparison signal
to the Channel 2 (or
center
plug-in) input connector. The reference signal
should
precede the comparison
signal in time. Use coaxial
cables or
probes which
have similar time-delay
characteristics to connect the signals
to the input connec
tors.
5. If the signals
are of opposite polarity invert the
Channel
2 (or center
plug-in)
display. (Signals may
be of
opposite polarity due
to
180°
phase difference; if so, take
this into account in
the final calculation.)
8. Move the waveforms to the center of the graticule
with the vertical
Position controls.
9. Turn the variable Seconds/Div control
until
one
cycle of
the reference signal (Channel
1, or left plug-in)
occupies exactly
eight divisions between the second and
tenth
vertical
lines of the graticule (see Fig. 1-8). Each
division
of
the graticule represents 45° of the
cycle (360°
4- 8 divisions=45°/division).
The
sweep rate can be
stated
in
terms of
degrees
as 45
“/division.
10.
Measure the horizontal difference
between cor
responding
points on the waveforms.
11.
Multiply the
measured distance (in divisions) by
45°/division
(sweep rate) to obtain the exact amount of
phase difference.
EXAMPLE:
Assume a horizontal difference of 0.6
division
with a
sweep rate of
45°/division as shown in Fig.
1-8.
Use
the formula:
Phase horizontal
Difference
=
difference
(divisions)
sweep
rate
X
(degrees/= (0.6) (45°)=27°
division)
The phase difference is
27°.
1-13
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......