Operating lnstructlons-5110
2.
Set
the
input coupling to ac and set the Volts/Div
switch
to
display
about 5 or 6 vertical divisions of the
waveform.
Check
that the variable Volts/Div control (red
knob)
is in the Cal
position.
EXAMPLE:
Assume a
peak-to-peak vertical deflection
of 4.6 divisions and a Volts/Div switch setting of 5 V.
Peak-to-peak
_
4.6 x 5
(Volts/Div
_
23
volts
(divisions) setting)
volts
3.
Adjust the
time-base triggering controls for a stable
display
and set
the Seconds/Div switch to display
several
cycles
of the
waveform.
4.
Turn the vertical
Position control so that the lower
portion of
the waveform coincides with one of the
graticule
lines below
the center horizontal line,
and the top
of the waveform is in the viewing
area. Move the display
with
the horizontal Position control so that one of the
upper peaks
is aligned with the center
vertical
reference
line (see
Fig.
1-3).
NOTE
If an attenuator probe is used that cannot change the
scale
factor readout (Volts/Div), multiply
the
right
side
of
the
above equation by the attenuation factor.
5. Measure the vertical deflection
from peak to peak
(divisions).
NOTE
This
technique
may
also be used to make
measurements
between two points on
the
waveform,
rather
than peak-to-peak.
Instantaneous
Voltage Measurement—DC
To
measure
the dc level at a given point on a waveform,
use the following
procedure:
1.
Set
the input coupling of the vertical plug-in unit to
Gnd
and position the trace to
the bottom line
of the
graticule (or other selected
reference line). If the voltage
to
be measured is negative with respect to ground,
position
the trace to the top line of the graticule. Do not
move
the vertical Postion control after this
reference has
been established.
6.
Multiply
the distance
(in
divisions) measured in step
5
by
the
Volts/Div switch setting. Also
include the
attenuation factor of the probe, if applicable.
Position
Io center
vertical line
Fig.
1-3.
Measuring peak-to-peak
voltage
of a waveform.
NOTE
To
measure a voltage level
with respect to a
voltage
other
than ground, make the following changes to
step
1:
Set the input
coupling
switch to dc and
apply
the
reference voltage
to the input connector, then
position
the
trace to the reference line.
2.
Connect the signal to the input connector. Set the
input
coupling
to dc (the ground reference can be
checked
at any time by setting the input coupling to Gnd).
3.
Set
the Volts/Div switch to display about 5 or 6
vertical divisions of the waveform.
Check that the
variable
Volts/Div
control (red
knob) is in
the Cal position.
Adjust
the
time-base triggering controls for a stable display.
4.
Measure
the
distance
in divisions between the
reference line and the point on the waveform at which the
de
level is to be measured. For example, in Fig. 1-4 the
measurement
is made between the
reference line and
point A.
1-8
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......