Specification
and Performance Check-5110
11. Check
Horizontal Bandwidth
a. Connect the
sine-wave generator to the amplifier
input, using
a 42-inch coaxial cable and 50 ohm termina
tion.
b.
Adjust the
sine-wave generator
controls
for a 6-
division
display at
a frequency of 50 kHz. Position the
display
between
the
second and eighth
vertical graticule
lines.
c.
Without
changing
the output amplitude, increase
the
sine-wave generator
frequency
until the displayed
amplitude is
reduced to 4.2 divisions.
d.
Check—the
generator for a
reading of at least
2
MHz.
e.
Disconnect the coaxial cable and termination from
the amplifier
input connector.
12. Check
X-Y
Phase Difference
a. Press the POWER
switch to
turn off the os
cilloscope.
b.
Remove
the time-base
unit from the vertical com
partment
and
install the second amplifier unit in the left
vertical compartment.
NOTE
Identical amplifier
units should
be
installed in the
oscilloscope.
c.
Connect the sine-wave generator through a
42-inch
coaxial
cable, 50
ohm termination, and atee
connector, to
an
amplifier
input.
Connect an 18-inch coaxial cable from
the tee connector
to the other amplifier input.
d.
Pull the oscilloscope
POWER switch to on.
e. Set
both amplifier units for a deflection factor of
1
volt/division and de input coupling.
f.
Set
the sine-wave
generator for a 100-kilohertz
output.
g.
Adjust the vertical and horizontal position controls
to
center the diagonal display, then adjust the sine-wave
generator
for
a display amplitude of 6 divisions vertically
and
horizontally.
h.
Check—
the opening of the diagonal-loop
display at
the
graticule center line is 0.07 division
or
less (measure
horizontally). This
indicates a phase difference of 1°
or
less between
the
vertical and horizontal
systems.
13.
Check
Z
Axis Amplifier
a. Press
the POWER switch to turn off the os
cilloscope.
b. Disconnect the coaxial cables, termination and tee
connector
between the
amplfiers
and sine-wave
generator.
c.
Remove
the
vertical
amplifier from the horizontal
compartment and
install
the time-base unit in that com
partment. Pull the oscilloscope
POWER switch to on.
d.
Set the time-base unit
for
auto, internal triggering at
a
sweep
rate of 20 µs/division and set the amplifier for a
deflection
factor of 2 V/division.
e.
Connect a
50 kHz sine-wave signal from the sine
wave
generator
through a 42-inch
coaxial
cable and a
tee
connector
to the amplifier input.
f.
Set the amplifier
and sine-wave generator controls
to obtain a calibrated 10
volt referencedisplay (5divisions
of display).
g.
Set the oscilloscope
INTENSITY control for a dim
display.
h.
Connect
the signal from the output of the
tee
connector
at the amplifier input,
to
the EXT
INTENSITY
INPUT connector on the front
panel.
i. Check—
the
top
of the waveform is intensified and
the bottom
portion is blanked out.
j.
Temporarily disconnect the coaxial cable at only the
EXT INTENSITY INPUT connector.
k.
Set the time-base
unit
for
a sweep
rate of
2
µs/division,
and increase the output frequency of the
sine-wave generator to 1
MHz.
l. Reconnect the coaxial cable
to the EXT INTENSITY
INPUT connector.
m.
Check—
for a noticeable effect of intensification in
the
top
portion of the displayed waveform and blanking in
the
bottom portion of the waveform.
n. Disconnect the
coaxial cables and tee connector
from
the amplifier and oscilloscope.
14.
Check
Calibrator Signal
a.
Connect
the
1X probe to the CH 1 input of the
amplifier.
Connect the probe tip
to the CALIBRATOR
loop.
2-9
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......