Adjustment-5110
12. Adjust
Horizontal Gain
a.
Press
the
POWER
switch to turn off the oscilloscope
and
interchange the amplifier and the time-base
units in
their respective compartments.
Pull
the POWER switch to
on.
b. Connect a 5-volt, 1-kilohertz square-wave signal of
standardized amplitude from the calibration generator to
the
amplifier input
connector, using a 42-inch
coaxial
cable.
NOTE
Use
an amplifier plug-in known to be accurately
calibrated, or
verify correct calibration by
applying a
known signal and meauring
the differential output at
pins
A7 and
B7 of
the plug-in connector. The
deflection factor at the output is
50 millivolts/division.
c.
Position
the
5-division display between the second
and seventh vertical graticule lines.
d.
Check—
the display for a horizontal deflection of 5
divisions,
±0.15 division
(±3%).
e.
Adjust—Hor
Gain R136
for exactly 5 divisions of
deflection. See
Fig.
3-4
for adjustment location.
f.
Disconnect the coaxial cable between the amplifier
and
the calibration generator.
13. Adjust
X-Y Phase
Difference
a. Press
the
POWER switch to turn off the os
cilloscope.
b. Remove the time-base
unit from the vertical com
partment and
install a second amplifier
in the left vertical
compartment.
NOTE
Identical
amplifier units should be
installed in the
oscilloscope.
c.
Connect
the sine-wave generator through
a
42-inch
coaxial
cable,
50 ohm termination, and a tee connector, to
an
amplifier input. Connect
an 18-inc coaxial cable from
the
tee connector
to the other
amplifier input.
d. Pull
the oscilloscope
POWER switch to on.
e. Set both amplifier units for
a deflection factor of
1
volt/division and de input coupling.
f.
Set
the sine-wave
generator for a 100-kilohertz
output.
g.
Adjust the vertical
and horizontal position controls
to center
the diagonal
display, then adjust the sine-wave
generator
for
a display amplitude of 6 divisions vertically
and
horizontally.
h.
Check
—
that
any opening of the diagonal-loop
display at the graticule center line is 0.07 division or less
(measured
horizontally). This
indicates
a phase difference
of 1°
or less
between the vertical and horizontal
systems.
i.
Adjust
—Phase
C116 for minimum loop opening (a
straight
line)
in the diagonal-loop display. See Fig. 3-4 for
adjustment location.
j. Press the POWER switch
to turn off the
oscilloscope.
k.
Disconnect
the coaxial cables,
termination and tee
connector
between the amplifiers and sine-wave
generator.
l. Remove
the vertical amplifier from
the horizontal
compartment
and install the time-base unit in that com
partment. Pull the oscilloscope
POWER switch toon. This
completes
the
Adjustment procedure
forthe oscilloscope.
3-7
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......