Operating lnstructions-5110
CHECKOUT
PROCEDURE
1.
For
the following
procedure, an amplifier plug-in
should
be in
one of the vertical (left or center) plug-in
compartments and
a time-base
plug-in should be in the
horizontal
(right) compartment.
2. Set
the POWER switch
to off (pushed
in) and
connect
the
oscilloscope
to a power source that
meets the
voltage and
frequency requirements of this instrument.
3. Turn the INTENSITY control counterclockwise and
pull
the
POWER switch out to turn the instrument on.
Initial
Control
Settings
Set the front-panel controls
as
follows:
NOTE
Titles for external
controls of
the
oscilloscope are
capitalized
in this procedure (e.g. INTENSITY,
POWER).
Trace
Alignment Adjustment
6.
If
a free-running trace
is not parallel with the
horizontal
graticule
lines,
set
the TRACE ROTATION
control (rear-panel
adjustment) as follows: Position the
trace
to the center horizontal
line and adjust the TRACE
ROTATION
control so that the trace is parallel with the
horizontal
graticule lines.
Calibration
Check
7.
Connects 1X probe,
or
a test lead from the amplifier
plug-in connector
to
the
CALIBRATOR loop.
8.
Set the time-base
unit triggering level for a stable
triggered display.
Adjust the vertical and horizontal
posi
tion controls
so that
the
display is centered vertically and
starts
at
the left
edge of the graticule.
9.
The
display should
be four divisions
in amplitude
with approximately 2.5 complete cycles over 10 divisions
(for
60-hertz
line
frequency) shown horizontally.
An
incorrect display
indicates that the Oscilloscope or plug
ins
need to
be recalibrated.
AMPLIFIER
PLUG-IN
Display
On
Position
Centered
Volts/Div
.1
Volts/Div
Cal
Fully
clockwise
Input coupling
de
TIME-BASE
PLUG-IN
Display
Chop
Position
Centered
Second s/Div
2 ms
Seconds/Div Cal
Fully
clockwise
Swp Mag
Off
Triggering
+ Slope, Auto Trig,
ac
Coupl
Triggering Source
Composite
Intensity
Adjustment
4.
Advance
the INTENSITY control until
the trace is at
the desired viewing level. Set the trace near the graticule
center
line.
Focus
Adjustment
5.
Adjust
the
FOCUS control fora sharp, well-defined
trace
over
the entire trace length.
Beam
Finder
Check
10.
Move
the
display off-screen with the vertical
position
control.
11.
Push
the BEAM FINDER
button and observe that
the
display compresses into the screen area. Reposition
the
display to screen center and release the BEAM
FINDER
button.
Disconnect the 1X probe or test
lead.
External
Intensity Input
12.
Connect a 5-volt, 1-kHz sine-wave or
square-wave
signal
to the EXT INTENSITY INPUT connector. Also, use
the signal
to externally trigger
the
time-base plug-in.
13. Slowly
rotate
the INTENSITY control
counterclockwise until the trace appears to be a series of
dimmed and
brightened
segments. The brightened
segments
correspond with
the
tops of
the signal input
waveform.
14. Disconnect the signal
setup.
This
completes the checkout procedure for the os
cilloscope.
Instrument operations not explained here, or
operations
that need further explanation, are discussed
under
General Operating
Information.
1-4
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......