VOLTAGE
AND WAVEFORM
CONDITIONS
WARNING
Dangerous potentials
exist at several
points throughout this instrument. When the
instrument
is
operated with the
covers removed, do not touch exposed connections or
components.
Some transistors have voltages present on their cases. Disconnect the
power source before replacing parts.
RECOMMENDED TEST
EQUIPMENT
Item
Specifications
Recommended
Type
Test oscilloscope system
Deflection factor,
1 mV to 50 V/div; input
impedance, 1 megohm;
frequency
response,
de to 2 MHz.
Probe: 10X
attenuation
probe compatible with
vertical
input.
Tektronix
5110, 5A13N, 5B10N
oscilloscope
system or
equiv. Use a
Tektronix P6108
or P6062A Probe.
Voltmeter
(Non-loading
digital
multimeter)
Range, 0
to 250
V; input
impedance,
10
megohms.
Tektronix DM 501
Digital
Multimeter with power module.
VOLTAGE
CONDITIONS
Voltage
measurements on this diagram were
made under the following conditions:
An
amplifier unit is installed
in the left vertical compartment (for power supply loading). INTENSITY control is set fully
ccw.
Voltmeter
common is connected to chassis ground.
WAVEFORM
CONDITIONS
OSCILLOSCOPE
UNDER TEST.
Install
an amplifier unit in the left vertical compartment and a time-base unit in the
horizontal
compartment..
Connect the CALIBRATOR output signal to the amplifier unit (set vertical input coupling to de
and volts/div
for a 2-division display). Set the time-base unit for internal auto-trigger, 2
ms/division sweep
rate.
TEST
OSCILLOSCOPE.
Set
the test oscilloscope triggering for auto mode with ac coupling from the internal
source and
set
vertical
input coupling to ac.
Connect a 10X Probe to the vertical input. Position the display as necessary.
NOTE
The
waveforms shown are actual waveform photographs taken with a Tektronix Oscilloscope Camera
System
and
Projected
Graticule. Vertical deflection factor shown on the waveform is the actual deflection factor from the probe tip.
Voltages
and waveforms on the diagrams are not absolute and may vary between instruments because of component
tolerances,
internal
calibration,
or front-panel
settings. Readouts are simulated in larger-than-normal type.
o
0
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......