Adjustment-5110
c.
Set
the time-base unit Seconds/Div switch to an
amplifier
position
or for the
slowest
sweep
rate.
d.
Turn
the
INTENSITY control slowly clockwise and
check for a visible spot display. Note that the spot
appears
when
the control
is between
its 10 and
11 o’clock position.
If the spot appears
when
the control is within the given
position,
proceed with step 7a.
e.
Set
the INTENSITY control to its 10 o'clock position.
f.
_ Adjust—Intensity Range R245,
using an insulated
screwdriver,
for
a very dim spot display.
See Fig. 3-3 for
adjustment
location.
7. Adjust
Astigmatism
a. Set
the INTENSITY control for a spot display at
normal
viewing level.
b. Turn the FOCUS control through its
range.
c.
Check—
for a spot
display that
is nearly round in
shape
when defocused, and well
defined when focused.
d.
Adjust—Astig R286
and FOCUS control together,
using
an insulated screwdriver, to obtain the best defini
tion
round-spot
display. See Fig. 3-3 for adjustment
location.
8.
Adjust
Trace Alignment
a.
Set
the
time-base unit for a sweep rate of
1
millisecond/division.
b. Set the INTENSITY control
for a trace of normal
brightness.
c.
Position the
horizontal trace over the
center
horizontal
graticule line.
d.
Check—thatthe trace
is parallel tothe
graticule line.
e.
Adjust—the TRACE ROTATION control
(rear-panel
screwdriver
adjustment)
to
align the
trace
horizontally.
9. Adjust Geometry
a.
Press the
POWER switch to
turn off the os
cilloscope.
b.
Interchange
the amplifier and time-base units in
their
respective
compartments.
Pull the
POWER switch to
on.
c.
Position
the vertical trace over the center vertical
graticule
line,
extending vertically above and below the
graticule
area, and set the
FOCUS and INTENSITY
controls for
a well-defined trace,
if necessary.
d.
Check
—that vertical bowing and tilt
of the trace
display
is less
than 0.1 division at
the
center line
and when
positioned
horizontally across the entire graticule area.
e.
Adjust—Geom R285 for minimum bowing and tilt
of
the trace
display at the left and right
edges of the graticule.
Adjustment
may
have to be compromised to obtain less
than
0.1 division bowing and tilt everywhere within the
graticule area.
See Fig. 3-3 for adjustment location.
f.
Press
the
POWER switch to turn offtheoscilloscope
and interchange the amplifier and time-base units back to
their usual compartments.
g.
Pull the
POWER
switch to on and check that
horizontal bowing
and tilt of the
trace display is less
than
0.1 division
at
the center line and when positioned
vertically
across the
entire graticule area.
Set
the equipment controls as
follows:
AMPLIFIER
PLUG-IN
Display
Position
Volts/Div
Volts/Div
Cal
Input Coupling
On
Centered
1
Fully
clockwise
de
TIME-BASE
PLUG-IN
Position
Seconds/Div
Seconds/Div Cal
Swp
Mag
Triggering
Triggering Source
Centered
1
ms
Fully
clockwise
Off
+
Slope,
Auto Trig,
ac Coupl
Composite
10. Adjust
Vertical Gain
a.
Connect a 5-volt, 1-kilohertz square-wave signal of
standardized amplitude from
the calibration generator to
the amplifier input,
using a 42-inch coaxial cable.
NOTE
Use
an amplifier
plug-in known to
be accurately
calibrated,
or
verify correct calibration
by applying a
known
signal and measuring the differential output
at
pins A7
and
B7 of the plug-in connector. The
deflection
factor at
the output is 50 millivolts/divi-
sion.
3-5
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......