VOLTAGE AND
WAVEFORM
CONDITIONS
WARNING
Dangerous
potentials
exist at several points throughout this instrument. When the
instrument
is operated with
the covers
removed, do not touch exposed connections or
components.
Some transistors
have voltages present on their cases. Disconnect the
power source
before
replacing parts.
RECOMMENDED
TEST EQUIPMENT
Item
Specifications
Recommended
Type
Test
oscilloscope system
Deflection factor, 1
mV to 50 V/div; input
impedance,
1
megohm; frequency
response,
de
to 2 MHz. Probe:
10X
attenuation probe compatible with
vertical
input.
Tektronix 5110, 5A13N, 5B10N
oscilloscope
system or
equiv.
Use a
Tektronix P6108
or
P6062B
Probe.
Voltmeter (Non-loading
Range, 0 to 250 V; input impedance,
Tektronix DM 501A Option
02 Digital
digital
multimeter)
10
megohms.
Multimeter
with
power module.
VOLTAGE
CONDITIONS
Voltage measurements
on this diagram were
made under the following conditions:
An amplifier
unit
is
installed in the left vertical compartment
(for
power supply loading). INTENSITY control is set fully ccw.
Voltmeter
common
is connected to chassis ground.
WAVEFORM
CONDITIONS
OSCILLOSCOPE UNDER TEST.
Install
an amplifier unit in the left vertical compartment and a time-base unit in the
horizontal
compartment. Connect the CALIBRATOR output signal to the
amplifier unit (set vertical input coupling to de
and
volts/div
for a 2-division display). Set the time-base unit for internal auto-trigger, 2
ms/division sweep rate.
TEST
OSCILLOSCOPE.
Set
the
test oscilloscope triggering for auto mode with ac coupling from the internal source and
set vertical input coupling
to ac. Connect
a 10X Probe to the vertical
input. Position the display as necessary.
NOTE
The
waveforms shown
are actual waveform photographs taken
with a Tektronix Oscilloscope Camera System and
Projected Graticule.
Vertical
deflection factor shown on the waveform is the actual deflection factor from the probe tip.
Voltages and
waveforms on the diagrams are not absolute and
may vary between instruments because of component
tolerances,
internal
calibration, or front-panel settings. Readouts are simulated in
larger-than-normal type.
G
G
500
mV
5 ms
—
-
G
G
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......