WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
TDI-Dynaload Division
Page 66
402828, Rev. B1
I1 Constant Current Pulsing Base Level
Description
This command is used to set the base level when pulse loading in the
Constant Current mode.
Command Syntax
I1 <NR2>
Example
I1 10.5
This will set the base Constant Current level to 10.5 Amps.
Query Syntax
I1?
Returned Parameters TEXT ON: <NR2> amps
TEXT OFF: <NR2>
I2 Constant Current Pulsing Peak Level
Description
This command is used to set the peak level relative to the base level
when pulse loading in the Constant Current mode.
Command Syntax
I2 <NR2>
Example
I2 25.0
This will set the peak level to 25 Amps above the base level.
Therefore if the base level setting is 10.5A, the peaks will be 35.5
Amps.
Query Syntax
I2?
Returned Parameters TEXT ON: <NR2> amps
TEXT OFF: <NR2>
ID? Identification
Description
This query returns the load’ s model identification string that
includes its rated Voltage, power and current in the format:
WCL <rated Voltage>-<rated power>-<rated current>
Command Syntax
NONE
Example
N/A
Query Syntax
ID?
Returned Parameters WCL <Voltage>-<current>-<power>