Page 63
DU Duty Cycle
Description
This command is used to set the Duty Cycle for square wave
pulsing. This defines the percentage of peak pulse time and will set
the Pulsing Base Time (T1) and Pulsing Peak Time (T2) accordingly
based upon the defined frequency.
Command Syntax
DU <NR2>
Example
DU 40.0
This will set the duty cycle to 40%. If the frequency was 100Hz
meaning a pulse period of 10,000µs, T2 will be set to 4000µs and T1
will be set to 6000µs.
Query Syntax
DU?
Returned Parameters TEXT ON: <NR2> %
TEXT OFF: <NR2>