CV — Constant Voltage Mode (Basic Mode)
The constant voltage mode can best be described as a shunt regulator or a zener diode. The load
will not conduct any current until the source voltage tries to exceed the voltage set point. Once
the source voltage is high enough the load will shunt current in order to regulate the voltage. The
regulating voltage is adjustable from full scale of the range selected to approximately zero. The
constant voltage mode is used to simulate a battery to a battery charger or for special applications,
such as a shunt regulator.
NOTE:
Never use the constant voltage mode for testing a constant voltage source. The regulators
of the two devices will buck each other trying to gain control of the voltage, which in
turn will lead to an unstable condition.
CP — Constant Power Mode (Basic Mode)
In constant power mode, the Dynaload will dissipate a set wattage anywhere up to the maximum
power rating of the unit. The Dynaload will automatically adjust the current level inversely in
response to a change in voltage. The constant power mode is
NOT affected by changes of the
voltage or current range.
CAUTION:
If the source voltage decays to zero volts the load will attempt to draw
infinite current.
Short Circuit
Due to the intended use and power levels involved, “Short Circuit” is not available on WCL or
WCS products.
MEM — Memory Recall Locations
The configuration of the load may be stored in 10 settable memory locations labeled 0 through 9.
To recall a memory location, press the MEM key on the front panel. Enter the numeric location
number, then press ENTER. When using the IEEE-488 bus the setup can be stored with the MS
command and recalled with the MR command.
NOTE:
Memory location “0” is reserved for initialization of the Dynaload.
Manual Adjust
The round knob above the MAN ADJ key is used to manually adjust all settable features within
the Dynaload. This knob can be activated by pressing the MAN ADJ key. The resolution of the
knob is set through the menu.
Remote Programming
The WCL series is analog programmable in all four modes of operation. The loading in a
particular voltage or current range is directly proportional to the 0-10 volt programming input.
For example: If the constant current mode is selected and the 200A range is selected, a