WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
TDI-Dynaload Division
Page 74
402828, Rev. B1
RNG Range
This command is used to set the load’ s Voltage and current range
settings. This is done by writing a value from 1 through 9 as
indicated below.
Range Number
Voltage
Current
1
High
High
2
Medium
High
3
Low
High
4
High
Medium
5
Medium
Medium
6
Low
Medium
7
High
Low
8
Medium
Low
9
Low
Low
Description
The Query form of this command returns actual operating range.
The text response to a query will be a string specifying the Voltage
and current ranges. The numeric response will be the range number.
Command Syntax
RNG <NR1>
Example
RNG 2
This will set the load to the medium Voltage and high current range.
Query Syntax
RNG?
Returned Parameters TEXT ON: <V range> VOLT, <I range> AMP
TEXT OFF: <NR1>