TDI-Dynaload Division
WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
402828, Rev. B1
Page 15
Volt, slow blow, 1-1/2 inch American version) or equivalent
OR GDC-3.15A-250V (3.15 Amp,
250 Volt, slow blow, 20 mm European version) or equivalent.
WCL488 AC Input Wiring Tips
A standard U.S. three-prong cord is provided with your Dynaload. The Input Voltage Selector is
mounted above of the three-prong AC connector and fuse (see Figure 11 and refer to the
WCL488
Input Voltage
Selector section that follows).
WARNING:
The power cord provides a chassis ground through a third conductor. Make
sure that your power outlet is of the 3-conductor type with the correct pin
connected to earth ground. Always Connect the AC cord first to your
Dynaload and THEN to the utility outlet.
WCL488 Slave Communication Connector
This DB25 “D” connector is located in the upper right corner of the rear panel and is used for
synchronized parallel Dynaload operation (see Figure 10). This connector is labeled “SLAVE
COMMUNICATION”.
NOTE:
There are no user signals or inputs in this connector.
WCL488 Slave Communication Connector Wiring
This 25-pin “D” connector is used to link the master unit to slave units. If you are purchasing
cables, be sure the PIN configuration is always 1 to 1 (not all cables are provided this way and a
crossover of PINS will cause erroneous operation).
WCL488 UV (Undervoltage Lockout) Switch
Referring to Figure 12, the UV switch has two positions, ON and OFF. This switch is described
in detail in the Operating Instructions chapter of this manual. Refer to the
Undervoltage Lockout
section later in this chapter.