WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
TDI-Dynaload Division
Page 78
402828, Rev. B1
SF Slew Fast
Set for fast slew range
Description
This command is used to set the slew rate limiter to the fast range. In
the fast range, full-scale rise and fall times can be programmed
between 10
µ
s and 4000
µ
s using the SR, S1 or S2 commands.
NOTE:
Maximum slew rate is highly dependent on factors including
power source, source to load wiring, and operating mode.
Command Syntax
SF
Example
SF
Sets the slew rate limiter to the fast range.
Query Syntax
NONE
Returned Parameters NONE
SR Slew Rate
Description
This command is used to set both the rise and fall time in
microseconds for a full-scale transition within the selected range.
Command Syntax
SR <NR2>
Example
SR 200
Sets the rise or fall time for a full-scale transition to 200
microseconds.
Query Syntax
SR?
Returned Parameters TEXT ON: <NR2> us
TEXT OFF: <NR2>