V2 Constant Voltage Pulsing Peak Level
Description
This command is used to set the peak level relative to the base level
when pulse loading in the Constant Voltage mode. In Constant
Voltage, peak refers to peak drive hence lower Voltage. Therefore
the peak level this command specifies is the Voltage to be subtracted
from the base level (V1).
Command Syntax
V2 <NR2>
Example
V2 24.3
This will set the peak level to 24.3 Volts below the base level.
Therefore if the base level setting is 75.0, the peaks will be 50.7
Volts.
Query Syntax
V2?
Returned Parameters TEXT ON: <NR2> volts
TEXT OFF: <NR2>
VER? Version
Description
This query returns the load’ s firmware version as a variable length
alphanumeric string.
Command Syntax
NONE
Example
N/A
Query Syntax
VER?
Returned Parameters <version string>
VL Voltage Limit
Description
This command is used to set the Voltage Limit trip level. The load
will disconnect from the device under test if the Voltage exceeds this
level.
Command Syntax
VL <NR2>
Example
VL 16.5
This will set the Voltage Limit to 16.5 Volts.
Query Syntax
VL?
Returned Parameters TEXT ON: <NR2> volts
TEXT OFF: <NR2>