Troubleshooting (Symptoms/Possible Causes/Explanations and Corrective Actions)
Symptom
Possible Cause
Explanation/Corrective Action
AC input fuse on the rear panel
has blown.
Check fuse and replace as
necessary.
AC power cord is loose or not
connected.
Secure AC power cord.
AC power button does
not function.
AC select on the rear panel is
not set properly.
2-Position Switch Models
Positions:
Low switch position 110 volts AC
High switch position 220 volts AC
4-Position Switch Models
Positions:
Lowest switch position 100 volts
AC
Next switch position 120 volts AC
Next switch position 200 volts AC
Highest switch position 240 volts
AC
Select the AC voltage you are
using.
Load powers up but
the red SYS fault LED
lights and the DC
LOAD LED will not
come on.
AC line voltage may be too
low.
The load detects low incoming AC
voltage (88% or less) and turns the
load off for self-protection.
Measure the AC input and correct
or change the AC switch position
as above.
Load powers up and
the “ SAT” LED lights
before a “ Load On” is
initiated.
There are one or more defective
components in the load.
The “ SAT” LED should never
come on during a load off
condition.
Send the unit in for service.
EN to S– jumper may be
missing.
Install missing jumper.
Load will not “ Load
On” with a green LED
indicator.
Bad EPO switch.
Verify EPO switch function.