Basic Usage
30
Operating as Constant Current Power Supply
Please verify that the settings for Over Voltage Protection
(OVP) circuit (page 41) and Over Current Protection (OCP)
circuit (page 43) are in effect.
Turn the POWER switch ON to startup the power supply.
The voltage measurement value will be displayed in the Number
Display (voltage) and the current measurement value will be
displayed in the Number Display (current). The lamp “V” and “A”
will light.
(If parameters for FUNCTION settings item 52 is set to 1,
PRESET value will be displayed)
Press the PRESET key
PRESET lamp will light.
Press the PRESET key again if you wish to discontinue and
return to the measurement display mode.
The voltage limit PRESET value will be displayed in the Number
Display (voltage), and the current PRESET value will be
displayed in the Number Display (current).
Turn the CURRENT dial to adjust the Constant Current Set Value.
(Turn the dial right to raise or left to lower the set value.)
If the parameter for FUNCTION item 50 is set to 0, press the
PRESET key again. The output current will be updated once the
PRESET mode has been exited.
If the parameter for FUNCTION item 50 is set to 1, promptly
change the set value using the CURRENT dial.
Turn the VOLTAGE dial to adjust the Voltage Limit Value
(Turn the dial right to raise or left to lower the set value.)
Continued on next page
When POWER is turned OFF for the top or
bottom Number Display, the Number Display
will be displayed in the same manner when
POWER is turned on again.
* The "kW" lamp will light when the power is
displayed.
MEMO
Press the CURRENT dial to display the current number
position. The selected number will be brightened. Press
to change position.
If all 4 places are brightened, numbers will change from
the lowest place.
MEMO
If the voltage limit value is set to a value lower
than the required load voltage, the voltage will be
restricted and the constant voltage will be used.
(CV lamp lights)
MEMO
* The above is the display cases (HX-S-030-200G2)
。
* The above is the display cases (HX-S-030-200G2)
。
Содержание HX-S-01000-12G2
Страница 165: ...Special Loads 165 This page is blank...
Страница 169: ...Maintenance 169 This page is blank...
Страница 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Страница 212: ...Specifications 212 2 HX S 030 400G2 F I...
Страница 213: ...Specifications 213 3 HX S 060 100G2 F I...
Страница 214: ...Specifications 214 4 HX S 060 200G2 F I...
Страница 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Страница 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Страница 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Страница 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Страница 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Страница 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Страница 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Страница 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Страница 223: ...Specifications 223...
Страница 239: ......