A-96.250.211 / 171121
AMI Sodium P
Troubleshooting
86
E011
Temp. shorted
– check wiring of temperature sensor
– check temperature sensor
E012
Temp. disconnected
– check wiring of temperature sensor
– check temperature sensor
E013
Case Temp. high
– check case/environment temperature
E014
Case Temp. low
– check case/environment temperature
E017
Control Timeout
– check control device or programming in
Installation, Relay
E018
Reagent empty
– check DIPA bottle
– if necessary refill diisopropylamine
E019
No Sample
– readjust the sample flow
– check if tubes are blocked
E020
pH low
– check process
E024
Input active
– information that the Input is active
– see If Fault Yes is programmed in
Menu
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2
E030
EEprom Frontend
– call service
E031
Calibration Recout
– call service
E032
Wrong Frontend
– call service
E033
Power-on
– none, normal status
E034
Power-down
– none, normal status
Error
Description
Corrective action